When an implant fractures, the device, the surgical record and the patient’s clinical course may each hold different parts of the explanation. A fracture alone does not establish a product defect, and a product investigation does not by itself establish whether clinical care contributed to the injury.
For manufacturers, the useful response is a coordinated evidence plan: preserve the device and its history, support a fair technical examination, and address clinical fault, product defect and causal contribution as separate questions. China’s Civil Code and the Supreme People’s Court’s medical-damage interpretation distinguish these issues and provide for evidence-based allocation where both product and clinical factors contribute.
Separate defect, clinical fault and causation
Organize the investigation around three questions. First, did the medical device fail to meet a legally relevant safety or performance requirement? Second, did a diagnostic or treatment act involve fault? Third, did either or both causes produce the claimed injury, and to what extent? The answers may overlap, but they require different records and expertise. Mechanical fracture analysis does not determine whether treatment met the applicable standard; a medical opinion does not replace examination of the device and its manufacturing history.
Articles 1202–1203 of the Civil Code address manufacturer and seller liability where a product defect causes damage, while Article 1218 addresses harm in diagnosis or treatment where the medical institution or staff is at fault. The Supreme People’s Court’s interpretation for medical-damage disputes identifies both device defect and clinical fault/cause contribution as questions that may require appraisal. It also states that where a defective medical product and negligent treatment together cause the same harm, joint liability to the patient may apply, with recourse between responsible parties determined by their respective causal contribution. The facts and proof in the particular case remain decisive.
Preserve the device without compromising the examination
On notice of a fracture or serious complaint, open a documented investigation promptly under the company’s quality, vigilance, complaint and legal-hold procedures. There is no general 72-hour civil-law deadline for the preservation steps below. Timing can matter in practice because the device may be cleaned, altered, discarded or returned to service, so assign an owner and agree on handling with the hospital and other custodians as soon as circumstances allow.
- Identify the item: record product name, model, serial or lot identifier, UDI where applicable, implant card details, packaging and the source from which each identifier was obtained.
- Document custody: photograph the device and packaging as received, record condition and transfers, identify every custodian, and preserve the original container and labels if available.
- Coordinate examination: invite relevant parties to agree on inspection steps, equipment, observers, sample allocation and recording. If testing could alter or destroy evidence, define the sequence and preserve images or other appropriate records first.
- Protect clinical evidence: through lawful channels, identify the records needed to understand implantation, device positioning, postoperative course, loading events, revision surgery and explant handling. Respect patient confidentiality and the custodian’s access rules.
Do not modify, polish, clean or destructively test the only specimen before documenting its received condition and agreeing how the examination will be conducted. Where no agreement is possible, keep the item secure and seek procedural advice on a neutral inspection or court-supervised preservation route. A company’s unilateral test may produce useful engineering information but can also create disputes about representativeness, handling or reproducibility.
Build the product-quality file around the specific failure
Link the returned device to the design and production records for its model and lot. Gather the applicable product technical requirements, approved design and manufacturing changes, process and inspection records, supplier information, nonconformities, complaints and any corrective or preventive action. Compare the fractured region and observed failure mode with the device’s design, intended use and documented acceptance criteria. Record both confirming and disconfirming observations; do not treat a single failure as proof that every device in the lot is defective or that the event is isolated.
The examination plan should identify questions an engineer or qualified technical expert can answer: the location and pattern of fracture, material or manufacturing anomalies, fatigue or overload indicators, evidence of damage after removal, and the limits of each test. Keep raw instrument outputs, calibration and method records, photographs, analyst identity, sample condition and every transformation used to reach a conclusion. State what cannot be determined because the sample is incomplete or its history is uncertain.
Keep the clinical and product inquiries coordinated
Prepare a chronology that joins, but does not conflate, device events and clinical events: manufacture and distribution, implantation, follow-up, reported symptoms, imaging, fracture discovery, revision and explant. Ask the treating institution to preserve relevant records through its proper process. Where the product issue and clinical-care issue are both disputed, request that any court-appointed medical-damage appraisal identify the specific questions, materials and expertise required. The Supreme People’s Court interpretation calls for authentic, complete and sufficient appraisal materials and a clear statement of the appraisal matters and requirements.
Medical-damage appraisal may address whether treatment involved fault, whether treatment caused harm and the degree of causal contribution, whether a medical product had a defect and whether that defect caused harm, and the injury consequences. These are distinct questions. A party may challenge an incomplete or unsuitable material set, an unclear question, or reasoning that exceeds the expert’s assigned scope. The court controls whether and how appraisal is ordered; a manufacturer should not assume it can require a particular separate “product appraisal” track.
- Clinical chronology: align operative notes, imaging dates, follow-up, symptoms and revision records.
- Device chronology: align product identifiers, lot history, shipment, complaint intake and return/explant custody.
- Question matrix: distinguish defect, clinical fault, causation and injury extent; identify the records and expert discipline relevant to each.
- Joint examination protocol: document agreed inspection, non-destructive steps, destructive testing conditions, observers and output retention.
- Limitations statement: record missing materials, uncertain history, alternative mechanisms and conclusions that the available evidence cannot support.
Run regulatory safety duties in parallel
Litigation strategy must not delay safety action. The Regulations on the Supervision and Administration of Medical Devices require a registrant or record-filer that discovers a device does not meet mandatory standards or registered product technical requirements, or has another defect, to stop production, notify relevant operators, users and consumers, recall devices already on the market, take remedial or destruction measures, keep records, disclose relevant information and report recall handling as required. That obligation is triggered by the regulatory facts; a complaint or lawsuit alone is not a substitute for the required defect assessment, and a pending dispute is not a reason to suppress a report or postpone a required action.
Route the event through the company’s adverse-event monitoring and risk-management process. Assess similar complaints, trend signals, affected configurations and whether additional investigation or field action is needed. Keep the safety assessment, investigation record and any recall decision linked but distinguishable from the legal position in the individual claim. Communications with the hospital and patient should be accurate, respectful and coordinated by appropriately trained personnel; avoid speculation about cause before the evidence is available.
What a defensible file should show
A useful case file lets another reviewer reconstruct what happened to the device, how it was examined, which clinical and product hypotheses were tested, what information was missing, and what safety action was taken. Preserve source records, not only summaries. Record disagreements rather than silently selecting one account. If the evidence supports more than one contributing cause, analyze that possibility directly instead of forcing a single-cause narrative.
This approach protects the integrity of the investigation and helps counsel present the issues in a form the court and its experts can answer. It cannot guarantee a particular appraisal result or liability allocation; the court determines the legal outcome from the evidence in the case.
Sources and limits
This article draws on the Civil Code of the People’s Republic of China, including Articles 1202–1203 and 1218; the Supreme People’s Court’s Interpretation on the Application of Law in the Trial of Medical-Damage Liability Disputes, including Articles 10–12 and 21–22; and the Regulations on the Supervision and Administration of Medical Devices, including Article 67. Appraisal procedure and safety-reporting obligations depend on the specific facts and current applicable requirements.